Cee® 1300CSXThermal Slide Debonder
The Brewer Science® Cee® 1300CSX thermal slide debonder enables high-temperature slide-off debonding of thinned compound semiconductor materials (GaAs, GaN, InP, and SiC) in a laboratory setting. This tool allows engineers to complete the final step of thinned full wafer processing in a developmental setting. Internal small-scale prototyping capability may be used to accelerate product development cycles and improve time to market for new compound semiconductor applications (high-power RF, LED, and solar).
Benefits
- Design permits in-house debonding of fully processed, proprietary ultrathin device wafers
- Enhanced data logging feature provides detailed process feedback and record keeping
- PC control allows virtually unlimited log storage
- Compact footprint permits installation flexibility
Tool Features
- Network connection hardware/software: RJ-45 Ethernet & USB port
- Precision lower platen z-position control (closed-loop control / feedback): ± 5 μm accuracy
- Specialized insertion and extraction end effectors with vacuum function (foot pedal control)
- Visual and audible light tree alarm
- Enhanced light curtain operation for seamless operation
- Continue abort override operation for "super user"
- Optional exhaust chamber sensor (PC-monitored)
- Optional hinged rear and side access panels
Utility Requirements & Dimensions
- Exhaust: 20-30 cfm at 1" W.G. (4" OD exhaust duct)
- Electrical: Voltage range 208-240 V, single phase, 50/60Hz, 3500 W
- Power Requirements: 18 amps
- Vacuum: -25" to -27" Hg (optimal vacuum: -27" Hg, (4.5 m3/h)
- Nitrogen or CDA: 100 psi, 1 cfm
- Optional Enclosure Purge: 3/8" push-to-connect (PTC tube) (20 psi)
- Dimensions: 49" W × 39.5" D × 52.5" H (125 cm W × 100 cm D × 133 cm H)
- Machine Weight: 415 lb (187 kg)
- Shipping Weight: 1,280 lb
Other Specifications
- Platen Maximum Temperature: 300°C
- Substrate Sizes (round): 2 in, 3 in, 100 mm, 125 mm, 150 mm, 200 mm
- Constant Force Mode: 0 to 100 lb (with maximum velocity limit of 100 mm/s)
- Logging: Critical force, distance, velocity, upper/lower platen vacuum, upper/lower platen temperature, z-position, and entire process duration time
- Excess Force Sensing: Failsafe error recovery